ARCS instrument Components

Overview

The ARCS instrument is a wide angle Fermi chopper spectrometer. As it selects the incident energy and uses time of flight to measure the final energy it is known as a direct geometry spectrometer, one of four at the SNS which are compared in Review of Scientific instruments 85, 045113 (2014) . An overview of the instrument is provided below. Followed by a table giving the main distances of the instrument. Then their is a brief discussion for the instrument components. A more complete description of the instrument is provided in Review of Scientific Instruments 83, 015114 (2012).

Overview of the ARCS instrument

Description Distance (m)
Moderator to T0 Chopper 8.77
Moderator to Fermi Chopper 11.61
Moderator to Monitor 1 11.831
Moderator to Sample 13.6
Sample to detector pixels in the Horizontal plane 3.0
Moderator to Monitor 2 18.5

Neutron Guide

The neutron guide reflects lower energy neutrons providing higher flux, at the expense of beam divergence, at the sample position. It starts in the Shutter and is elliptically shaped to focus the beam to the sample position. The last section of guide is inside the sample chamber and is removable for when larger pieces of sample environment equipment are used.

T0 chopper

The purpose of the T0 chopper, which is located 8.77 m from the moderator, is to suppress the prompt pulse of fast neutrons produced when the proton beam strikes the target. This suppression is accomplished by having an ~0.20-m-thick piece of the alloy inconel in the beam when the proton pulse hits the target. This piece of inconel must be out of the beam in sufficient time for the 0.01 - 1 eV neutrons to pass. It only turns counter clockwise and can operate at rotational speeds between 30 Hz and 180 Hz in multiples of 30 Hz. The following table provides guidance on the choice of the T0 chopper base on your choice of incident energy. Note for energies larger than 1eV, the T0 chopper is not fully closed when the prompt pulse hits the target. Thus the background is somewhat elevated for these incident energies. Please consult with instrument staff about the appropriate T0 chopper frequency for your experiment.

Ei (meV) T0nu (Hz)
8 30
10-15 30, 60
20-30 30, 60, 90
40 60, 90, 120
50-70 60, 90, 120, 150
80-100 60, 90, 120, 150, 180
125-250 90, 120, 150, 180
300-500 120, 150, 180
600 150, 180
800-10000 180

Fermi Choppers

The Fermi chopper is the primary energy selection device on ARCS. It consists of a series of closely spaced neutron-absorbing blades (slit package) held together by a rotor that spins about a vertical axis in the path of the beam. The slit package is curved to be optimized for a specific Ei at a specific frequency. All slit packages are 100 mm in length. Each Fermi Chopper can be spun from 0 to 600 Hz in increments of 60 Hz. Two Fermi complete choppers are mounted on a translation table so that either can be chosen independently. The are installed on ARCS 11.61 m from the moderator. The table below provides the parameters of the Fermi choppers. Only one Fermi chopper may be spinning at full speed at a time. When decreasing the rotation speed of the chopper, use increments of 60 Hz down to 360 Hz. Otherwise the chopper will fault and needs to be reset by the instrument staff or Hall coordinators. Your local contact can assist you in choosing the appropriate chopper and rotational speed for your experiment.

Slit Package Description

Name Translation Table Position # Chopper number Optimum Ei @ 600 Hz (meV) # of slats Channel Width (mm) radius of curvature (m)
100-1.5-AST 3 2 100 32 1.52 0.580
700-1.5-AST 1 1 700 32 1.52 1.535
700-0.5-AST     700 72 0.51 1.535

Slits

There are two sets of beam defining slits on ARCS. The most upstream is after the Fermi chopper but before the Monitor. The second set of slits is located in one of two positions. When smaller sample environment equipment is used, it is located 20 cm upstream of the sample position. If larger sample environment is used it is located just inside the wall of the sample vessel.

Each set of slits has four independently operating blades. Each blade will fully cover the beam at its full extent of travel. Each blade is made of 99% isotopically enriched hot pressed 10B4C blades (74 – 77% B, 23 – 26% C). The slits are typically controlled by a set of virtual motors that provide a center and gap of the beam in the horizontal and vertical directions. Typically only the second set of slits is adjusted to limit the beam on the sample.

The default setting for slit 1 is for them to be centered at 0 mm in both directions. The vertical gap is 60 mm and the horizontal gap is 56 mm.

For slit 2 each blade has a full travel of 50 mm. 25mm is completely out of the beam and -25mm is completely in the beam.

Attenuators

Two attenuator plates, located between the Fermi chopper and the upstream slits, may be independently placed in the incident beam. Each plate is 1 mm thick and is 2% borated aluminum.

Oscillating Radial Collimator

Inside the sample chamber there is an oscillating radial collimator on a vertical translation table. For detectors with a scattering angle above ~30o The radial collimator filters out scattering that is from components that are beyond ~20mm from the center of the sample The collimator blades are spaced at 1.6o and there is a 12.8o gap around the straight through beam. This gap reduces potential scattering from the blades in a region where the collimator is ineffectual. The oscillation is 1.6o in a period of 21s. Details of the Collimator are provided in: (M. B. Stone et al. Rev. Sci. Instrum. 85, 085101 (2014) and M. B. Stone et al. EPJ Conf. 83, 03014) For sample environments that contain their own vacuum and/or several heat shields, this can significantly reduce the background. However for experiments using very low background sample environment equipment, this is not necessary and should be lowered out of the beam. For the largest Sample environment the Radial Collimator can not be used.

The radial collimator is most efficient near 90o scattering and does nothing at the lowest scattering angles. Therefore it works better for studies that do not analyze in the low Q region. Discuss with the instrument staff if the radial collimator is appropriate for your specific experiment.

Sample Environment

Details of sample environment mounting are provided on the sample environment website

More instrument specific sample environment information is provided here

Vacuum Chambers

The sample environment is positioned in a vacuum chamber that is contiguous with a detector vacuum chamber for the purpose of minimizing extraneous material between the sample and detectors.The chambers are operated below 1x10-5 Torr. Both chambers are lined with B4C neutron absorber such the the detectors only see the sample space or the absorber. The chambers have additional shielding on the outside to minimize the time independent background. A gate valve isolates the two chambers during sample changes to minimize the pumping time during such a change.

Detectors

The detector array on ARCS is an assembly of 1.0 m long by 25 mm diameter Linear Position Sensitive Detectors (LPSDs). The array has 920 detectors grouped into packs of 8 and located on a vertical cylinder with a radius of ~ 3.0 m (which is the distance between the sample position and the detectors in the horizontal plane). The detectors cover an angular range of the cylinder from -28o to 135o in the horizontal.

To avoid cross talk between detectors, absorbing baffles are placed around the tank. Nevertheless since ARCS has such a large angular coverage, all tube to tube scattering can not be eliminated. One should be wary of features that appear near 75% of Ei this can be examined by looking at the instrument view in Mantid to see if extra counts are observed in this energy transfer range in the lowest and highest angle detectors.

The LPSDs are filled with 3He at a pressure of 10 Atmospheres (1.0 MPa). To reduce background, there are no windows between the sample and the detectors; the detectors are located inside a detector chamber that is evacuated to a high vacuum (below 10-6 atm) and is contiguous with the sample vessel. An incoming neutron is converted through the nuclear reaction n0 + 3He → 3H + 1H + 0.764 MeV into charged particles tritium (T or 3H) and protium (p or 1H) which then are detected by creating a charge cloud in the stopping surrounding gas. The electrons from the ionized gas are collected at an anode wire running down the center of the detector tube. This wire is at 1870 V above ground and has a high resistance so the proportion of charge seen at each end allows one to determine the position of the neutron detection event. The length of the detectors is divided into 128 pixels of ~1. mm length by the electronics. Each pixel subtends an angle of ~0.485° perpendicular to the length of the tube and ~0.125° along the length of the tube. The pixels are often binned together during reduction. Discuss the current binning with your Local Contact. Each pixel has a timing resolution of 1 µs and saturates at no less than 70,000 n/s. After saturation a tube is ready for measurement within 10 µs.

There are two labeling conventions for the detector banks the following table maps the bank number and the bank name.

Bank number Bank name
1 B1
2 B2
3 B3
4 B4
5 B5
6 B6
7 B7
8 B8
9 B9
10 B10
11 B11
12 B12
13 B13
14 B14
15 B15
16 B16
17 B17
18 B18
19 B19
20 B20
21 B21
22 B22
23 B23
24 B24
25 B25
26 B26
27 B27
28 B28
29 B29
30 B30
31 B31
32 B32
33 B33
34 B34
35 B35
36 B36
37 B37
38 B38
39 M1
40 M2
41 M3
42 M4
43 M5
44 M6
45 M7
46 M8
47 M9
48 M10
49 M11
50 M12
51 M13
52 M14
53 M15
54 M16
55 M17
56 M18
57 M19
58 M20
59 M21
60 M22
61 M23
62 M24
63 M25
64 M26
65 M27
66 M28
67 M29
68 M30
69 M31
70 M32A
71 M32B
72 M33
73 M34
74 M35
75 M36
76 M37
77 M38
78 T1
79 T2
80 T3
81 T4
82 T5
83 T6
84 T7
85 T8
86 T9
87 T10
88 T11
89 T12
90 T13
91 T14
92 T15
93 T16
94 T17
95 T18
96 T19
97 T20
98 T21
99 T22
100 T23
101 T24
102 T25
103 T26
104 T27
105 T28
106 T29
107 T30
108 T31
109 T32
110 T33
111 T34
112 T35
113 T36
114 T37
115 T38

Beam Monitors

There are two neutron beam monitors located along the direct beam, one located just downstream of the Fermi chopper and variable aperture and a second located farther downstream of the sample near the beamstop. These two monitors are primarily used to determine the speed of the incident neutrons. To detect the neutrons, the beam monitors use a low pressure of 3He

Related Links